CT Lab

Chapter 21: SPECT and the Physics of Collimation

Nuclear medicine. How single photons are given a direction. The resolution–sensitivity trade-off of the parallel-hole collimator, through the system-resolution formula and a simulation.

In the previous chapter we put PET and SPECT side by side and saw how PET fixes direction through coincidence detection. This chapter narrows in on SPECT. Reconstructing the activity distribution that SPECT measures rides directly on the tools we built for CT (FBP, or the OSEM of Chapter 8). The reason SPECT still earns its own chapter is that its image quality is decided almost entirely by a single sheet of lead: the collimator. Here the basic trade-off of imaging-system design (resolution and sensitivity cannot both be had) appears in its simplest form.

Why a collimator is needed

PET counted the pair of 511 keV photons flying in exactly opposite directions and fixed the direction from the line (the LOR) joining the two detection points. The tracer a SPECT study uses (such as 99m^{99m}Tc) emits a single gamma ray. With no partner photon, one detection alone cannot tell us which direction it came from.

So SPECT places a lead collimator in front of the detector: a plate perforated by countless fine parallel holes that passes only the gamma rays aligned with the holes and absorbs the obliquely incident ones in the lead walls (the septa). Direction is selected not by electronic coincidence but by mechanical absorption. This approach (throwing away most of the gamma rays to obtain direction) is the main reason SPECT's sensitivity is more than an order of magnitude below PET's.

Direction selectiondetectorPbsourceresolution ↔ sensitivityshort holeswide acceptance = high sensitivity, low resolutionlong holesnarrow acceptance = low sensitivity, high resolution

The two faces of a parallel-hole collimator. Left: only gamma rays aligned with the holes reach the detector; oblique ones are absorbed in the lead septa (the principle of direction selection). Right: lengthening the holes narrows the acceptance angle, raising resolution but reducing the number of photons passed — sensitivity (the resolution–sensitivity trade-off).

Resolution worsens with distance

The geometric resolution of a parallel-hole collimator is set by how far the source is from the plate. With hole diameter dd, effective hole length LeffL_\text{eff}, and source-to-collimator distance zz, the blur of a point source (full width at half maximum) is approximately

Rcol  =  d(Leff+z)LeffR_\text{col} \;=\; \frac{d\,(L_\text{eff} + z)}{L_\text{eff}}

At z=0z=0 (flush against the plate) Rcol=dR_\text{col}=d, the hole diameter itself. The farther the source, the more the image blurs by the angular spread a single hole accepts, and RcolR_\text{col} grows linearly with zz. Unlike the focal blur of CT this is not magnification but plain degradation, which is why SPECT keeps the detector as close to the body as possible while it rotates.

Getting close to the patient is about resolution

The reason a SPECT gamma camera orbits along the body contour on a non-circular path is to keep zz small and preserve RcolR_\text{col}. Just 5 cm of extra distance visibly worsens the resolution.

The detector itself also has an intrinsic resolution RintR_\text{int} (from the crystal and position logic), and the resolution of the whole system combines the two in quadrature:

Rsys  =  Rcol2+Rint2R_\text{sys} \;=\; \sqrt{R_\text{col}^2 + R_\text{int}^2}

At short range RintR_\text{int} still matters, but a little farther out RcolR_\text{col} dominates. This is why SPECT image quality is said to be decided by the collimator.

Raising resolution lowers sensitivity

So why not just make the holes thinner and longer to shrink RcolR_\text{col} without limit? Because the thinner and longer the hole, the narrower the solid angle of gamma rays it accepts, and the fewer photons it detects (that is, sensitivity, or geometric efficiency, falls). The efficiency is approximately

g    (KdLeff) ⁣2 ⁣(dd+t) ⁣2g \;\approx\; \left(\frac{K\,d}{L_\text{eff}}\right)^{\!2}\!\left(\frac{d}{d+t}\right)^{\!2}

where tt is the septal thickness and KK a hole-shape constant (about 0.26 for hexagonal holes). Efficiency scales as (d/Leff)2(d/L_\text{eff})^2, so doubling LeffL_\text{eff} to improve resolution drops sensitivity to a quarter. Efficiency is almost independent of distance zz (farther away fewer rays reach a given hole, but more holes come into view and the two cancel). Only resolution degrades with distance; sensitivity stays put. That is the character of parallel-hole geometry.

Lower sensitivity means fewer counts collected in the same time. Radioactive counting obeys Poisson statistics, so fewer counts make a grainier image. A high-resolution collimator is therefore "sharp but noisy," a high-sensitivity one "smooth but blurred." Clinically, low-energy high-resolution (LEHR) and high-sensitivity collimators are chosen to match the task.

Simulation: moving sharpness and noise together

A phantom with hot spots and a pair of closely spaced line sources is blurred by the system resolution and given Poisson noise whose count level is proportional to sensitivity. Lengthening the hole LeffL_\text{eff} shrinks RsysR_\text{sys} until the two line sources separate, but the relative sensitivity drops and the image grows grainy. Increasing the source distance zz lets you confirm the other half: resolution alone degrades while sensitivity holds. The plot on the right shows how the current collimator's resolution worsens linearly with distance.

True distribution

WL 0.500 / WW 1.20Drag to adjust WL/WW

SPECT image (blur + noise)

WL 0.500 / WW 1.20Drag to adjust WL/WW

Resolution vs distance

050100150200250051015202530Source distance z (mm)System resolution (mm)
System resolution R_sys6.5 mm
Relative sensitivity100%

A phantom of hot spots and a closely spaced pair of line sources is blurred by the system resolution and given Poisson noise whose count level tracks sensitivity. Lengthening Leff raises resolution until the two line sources separate, but sensitivity falls and the image grows grainy. Increasing distance z degrades resolution linearly while sensitivity holds. High resolution is sharp but noisy, high sensitivity smooth but blurred. This is the core SPECT trade-off.

The reconstruction itself is the same as CT

Once the collimator has selected direction, projections at each angle (a sinogram) are in hand and reconstruction is no different from CT. In practice, even for this modality where attenuation and scatter matter, the mainstay is the OSEM of Chapter 8 with attenuation and resolution correction folded in. What makes SPECT special is the entrance to the measurement (how a single photon is given a direction), and once past that, the skeleton of the inverse problem is shared. That is the view this textbook has kept returning to.

References

  • Cherry SR, Sorenson JA, Phelps ME. Physics in Nuclear Medicine, 4th ed. Elsevier (2012) — standard derivation of collimator resolution and sensitivity.
  • Anger HO. Scintillation Camera. Review of Scientific Instruments 29, 27–33 (1958) — the original gamma camera.
  • Gunter DL. Collimator characteristics and design. In Nuclear Medicine, Henkin RE et al. (eds.), Mosby (1996).
  • Bruyant PP. Analytic and Iterative Reconstruction Algorithms in SPECT. Journal of Nuclear Medicine 43, 1343–1358 (2002).

On this page